OTT PLS-C
Pressure probe with conductivity measurement cell
The OTT PLS-C measures water conductivity, level, and temperature in both surface and groundwater applications with a robust ceramic pressure cell and 4-electrode conductivity cell. Its extremely low power consumption makes it ideal for long-term deployment, particularly at solar powered measuring stations for ground-and surface waters.
- Eaux de surface, Débitmétrie
- Vented pressure cell and 4-electrode graphite conductivity cell
- Water level, Pressure, Conductivity, Temperature
- Water level, temperature, and conductivity measurement - for use with external data logger
- 1
- SDI-12 or RS-485 (using SDI-12)
Outputs for conductivity include specific conductivity, salinity, and total dissolved solids
- The robust ceramic pressure cell offers industry-leading accuracy and does not deform over time like membrane technology, providing long-term measurement stability
- Rugged design: Ceramic pressure cell resistant to physical force (4 x burst pressure) and enclosure made of high-quality saltwater resistant steel for use in coastal environments
- Vented pressure probe, automatically compensates for changes in barometric pressure
- Robust cable with Kevlar core for length stabilization and capillary tube for venting
- Built-in microcontroller – compensates for temperature effects and applies correction values for gravitational acceleration or water density
- Simple setup and connection to external dataloggers using SDI-12
Water level, temperature, and conductivity measurement of:
- Groundwater wells
- Streams, rivers, channels, and canals
- Reservoirs, lakes, and wetlands
Ideal for monitoring:
- Continuous monitoring networks
- Short-term or long-term projects
- Salt water intrusion
- Aquifer storage and recovery
- Estuaries and wetlands
|  |  |  | 
| 
			OTT PLS-C - CE Konformitätserklärung | Declaration of Conformity | Declaration de Conformité
		 |  | 
                                
                                        
                                            
                                                    Download PDF
                                                
                                        
                                    
                                
	
		
                                
                                        
                                            
                                                    Download PDF
                                                
                                        
                                    
                            
                        
                        
                        
                    | Output parameters | Water level/pressure, temperature, specific conductivity, salinity, TDS | 
| Water level measurement (pressure) | |
| Pressure sensor | ceramic, temperature-compensated | 
| Measuring range | 0 ... 4 m, 0 ... 10 m, 0 ... 20 m, 0 ... 40 m, 0 ... 100 m water col. | 
| Resolution | 0.001 m; 0.1 cm; 0.01 ft; 0.1 mbar; 0.001 psi | 
| Accuracy (linearity + hysteresis) | ≤ ± 0.05 % FS | 
| Long-term stability (linearity + hysteresis) | ≤ ± 0.1 %/a FS | 
| Zerodrift | ≤ ± 0.1 % FS | 
| Pressure sensor capability to withstand overloads without permanent mechanical damage | ≥ 4 x measuring range | 
| Temperature-compensated operating range | -5 °C ... +45 °C (ice free) | 
| Units | m, cm, ft, mbar, psi | 
| Temperature measurement | |
| Sensor | NTC | 
| Measuring range | -25 °C ... +70 °C (ice free) | 
| Calibrated range | +5 °C ... 45 °C | 
| Resolution | 0.01 °C | 
| Accuracy | ±0.1 °C | 
| Units | °C, °F | 
| Conductivity measurement | |
| Sensor | 4 graphite electrodes | 
| Calibrated range | +5 °C ... 45 °C | 
| Measuring range 5 ... 2.000 µS/cm | |
| Resolution | 1µS/cm | 
| Accuracy | ±1 µS/cm or ± 0.5 % of measured value (whichever is higher) | 
| Unit | µS/cm | 
| Measuring range 0.1 ... 100 mS/cm | |
| Resolution | 0.01 mS/cm | 
| Accuracy | ± 0.01 mS/cm or ±1.5 % of measured value (whichever is higher) | 
| Unit | mS/cm | 
| Options | |
| Temperature compensation, conductivity | freshwater, saltwater, standard method 2510, ISO 7888/EN27888 | 
| Salinity calculation | Standard method or USGS 2311 | 
| Electrical data | |
| Supply voltage | 6 ... 27 V DC, typically 12/24 V DC | 
| Power consumption | |
| SDI-12 sleep-mode | <30 µAV | 
| SDI-12 active-mode | <32 mA | 
| Interfaces | SDI-12, RS-485 (SDI-12 protocol) | 
| Mechanical data | |
| Dimensions | |
| Probe (Ø x h) | 317 mm x 22 mm | 
| Cable length | SDI-12: 1 ... 100 m RS-485: 1 ... 1000 m | 
| Material | |
| Housing material probe | POM, stainless steel (DIN 1.4539, 904 L ), resistant to sea water | 
| Cable jacket | PUR | 
| Weight | |
| Probe | approx. 0.43 kg | 
| Probe cable | approx. 82 g/m | 
| Ambient conditions | |
| Storage temperature | - 40 °C ... + 85 °C | 
| Type of protection | Probe: IP 68 | 
| EMC limits | EG 2004/108/EG, EN 61326-1:2013 | 

 
							 
							 
							 
							 
							 
							 
							 
					
				 
					
				 
					
				 
					
				 
					
				 
					
				