Hydrolab MS /DS - Series 5

Technical Data


Sensors

Measurement Range Accuracy Resolution
Disolved oxigen(Luminescence) LDO 0-60 mg/l
*exceeds max. natural concentration

±0.1 mg/l
for values ≤ 8 mg/l
±0.2 mg/l
for values > 8 mg/l and ≤ 20 mg/l
± 10% reading
for values >20mg/l

0.01 mg/l

Disolved oxigen
(Clark cell)

0-50 mg/l ±0.2 mg/l
for values ≤ 20 mg/l ±0.6 mg
for values > 20 mg/l
0.01 mg/l
Conductivity sensor 0-100 mS/cm ± (0.5 % of reading
+0.001 mS/cm)

Salinity 0-70 ppt ±0.2 ppt 0.01 ppt
pH sensor 0-14 pH units ±0.2 units 0.01 units
Self cleaning
Turbidity sensor
0-3000 NTU <100 NTU: ±1 %
<400 NTU: ±3 %
<3000 NTU: ±5 %
≤400 NTU: 0.1 NTU
>400 NTU:
1 NTU
Turbidity sensor,
4 beam
0-1000 NTU ± (0,5 % of reading
+ 1 NTU)
≤ 100 NTU: 0.1 NTU
> 100 NTU: 1 NTU
Depth sensor 0-10 m (ref. pressure)
0-25 m
0-100 m
0-200 m
±0.003 m
±0.05 m
±0.05 m
±0.1 m
0.001 m
0.01 m
0.01 m
0.1 m
Chlorophyll a Dyn. measurem. range
low sensitivity: 0.03-500 µg/l
med. sensitivity:
0.03-50 µg/l
high sensitivity: 0.03-5 µg/l
±3 % signal level equivalents
of 1ppb rhodamine WT dye or higher using a rhodamine sensor
0.01 µg/l
Cyanobacteria Dyn. measurem. range
low sensitivity:
150-2,000,000 cells/ml
med. sensitivity:
150-200,000 cells/ml
high sensitivity:
150-20,000 cells/ml
±3 % signal level equivalents
of 1ppb rhodamine WT dye or higher using a rhodamine sensor
20 cells/ml
Ammonium* 0-100 mg/l-N ±5 % of reading;
at least ±2 mg/l-N
0.01 mg/l-N
Nitrate* 0-100 mg/l-N ±5 % of reading;
at least ±2 mg/l-N
0.01 mg/l-N

Chloride*

*max depth 15 m

0.5-18,000 mg/l ±5 % of reading;
at least ±2 mg/l
4 digits
TDG 533-1866 mbar ±1.9 mbar 1.3 mbar
ORP -999-999 mV ±20 mV 1 mV
Rhodamine WT Dyn. measurem. range
low sensitivity: 0.04-1,000 ppb
med. sensitivity: 0.04-100 ppb
high sensitivity:
0.04-10 ppb
±3 % signal level equivalents
of 1ppb rhodamine WT dye or higher using a rhodamine sensor
0.01 ppb
Ambient light 0-10,000 µmol s-1m-2 ±5 % of reading

1 µmol s-1m-2
Temperature –5 ... +50 °C ±0.1 °C 0.01 °C